NEW DEVICES Percutaneous Pulmonary Valve Implantation: The Melody® Valve The right ventricle (RV) to main pulmonary artery (PA) conduits that are used to reconstruct the right ventricular outflow tract in congenital heart diseases are prone to develop valvular incompetence and/or obstruction with time. These pejorative evolutions are associated with exercise intolerance, arrhythmias, and an increased risk of sudden death 9 and require multiple open-heart surgeries to replace the pulmonary valve.
Percutaneous pulmonary valve implantation was introduced as a new treatment option in patients with dysfunctional conduits.10,11 This technological breakthrough aims at prolonging the lifespan of RV to PA conduits and thus postponing open-heart surgery. The trans-catheter pulmonary valve (Melody®; Medtronic, Minneapolis, MN) is composed of a bovine jugular venous valve and a balloon-expandable stent made of a platinum-iridium wire. The current largely accepted indications for the use of a Melody® valve include12:
- A significant RVOT obstruction, defined as RV pressures > 2/3 of systolic blood pressure (SBP) with symptoms, or > 3/4 of SBP without symptoms
- A severe pulmonary regurgitation and RV dysfunction or RV dilatation or impaired exercise capacity
- Along with morphological criteria allowing a safe implantation site: RVOT dimensions < 22 × 22 mm and > 14 × 14 mm
The implantation procedure is standardized and safe, with a procedural mortality < 0.2%. The main complication to avoid during the implantation is coronary compression or occlusion, which can be evaluated by a pre-implantation balloon inflation in the RVOT. Other complications during implantation are the dislodgement of the device when implanted in distensible and dilated RVOTs and the risk of homograft rupture. Valve implantation significantly reduces the gradient across the outflow tract, RV pressures, and the pulmonary regurgitation,13 and significantly improves symptoms. Lurz et al.13 demonstrated that during a median follow-up of 28 months freedom from reoperation was 93% (±2), 86% (±3), 84% (±4), and 70% (±13), at 10, 30, 50, and 70 months, respectively. The main complications of the new generation of this innovative technology are late endocarditis and stent fractures in 20%.14 These stent fractures are silent in the majority of cases and are treated in symptomatic patients with RVOT stenosis by a Melody® valve-in-valve implantation. Pulmonary valve implantation is becoming the standard procedure in the treatment of dysfunctional conduits. It has been accepted by the regulatory agencies for distribution and use in Europe in 2006 and US Food and Drug Administration in 2010. By prolonging the lifespan of RV–PA surgically placed conduits, this innovative technology has reduced the number of multiple open heart operations in children and young adults with congenital heart disease, and may improve their life expectancy and life quality. As with all evolving new technologies, new generations of Melody valves were created in order to reduce current limitations and extend the spectrum of potential clinical indications. Improvements brought to the Melody® valve during the last few years of development or currently in progress include:
- Device design improvements
- Delivery system improvements
- Patient selection improvements using three-dimensional echography and MRI
- Dilatation with high-pressure balloon after implantation (to reduce residual gradients)
- Stent-in-stent implantation
- Structural improvements to extend this technology to patients with native, dilated, and distensible RVOT
These principles of percutaneous valve implantation are currently investigated in other off-label clinical settings. For instance, valves developed for trans-catheter replacement of the aortic valve were implanted in the pulmonary position for patients with larger annulus.15 A new device allowing the implantation of a pulmonary valve in a RVOT previously repaired with a transannular patch is also currently investigated but not published yet. Tissue-Engineered Valved Conduits: Decellularized Scaffolds, Polymer Scaffolds, and in Situ Regeneration The ideal RV–PA conduit for reconstruction of the RVOT still does not exist.
Cryopreserved homografts need a revision surgery in 36% and 90% of cases after 10 and 15 years, respectively.16–18 Hancock conduits need to be replaced after 10 years in 68% of cases, and 50% of Carpentier–Edwards Perimount® (Edwards Life-sciences, Irvine, CA, USA) valves (bioprosthetic stented valve made of bovine pericardium) implanted in children also have to be replaced after 5 years.19 Children younger than 2 years old operated with a Contegra® Medtronic conduit have to undergo a revision surgery in 67% of cases for failure.20 The reoperations needed to replace a failing conduit carry a significant risk of mortality (1%–3%) and morbidity: hemorrhagic syndrome, cerebral vascular accident, coronary damage, cardiac rhythm alterations, or infection. These complications translate into prolonged hospitalization and attendant costs. Surgical techniques have improved during the last three decades, but conduit failure and morbidity and mortality still occur (Table 1). Autologous pericardial valved conduits for RVOT reconstruction showed superb properties, but data for long-term follow-up are lacking.21  | Table 1 Current Surgical Valved Conduits to Replace the Right Ventricular Outflow Tract. |
As a consequence of the limited treatment options and the requirements for repeat surgery in children as they grow, new alternatives were investigated to reconstruct the RVOT. The advanced-therapy medicinal products (ATMPs) derived from the concept of regenerative medicine are presently seen as one of the main routes to reduce the above-mentioned risks, with the exception of organ transplantation. On the basis of these issues, the search for the ideal material to replace the RVOT started. The in vitro creation of autologous and living substitute materials by tissue engineering is based on the essential need for growth potential of materials to be used for surgical correction of congenital cardiac defects. In the last 15 years, different tissue-engineered materials have been proposed to replace the RVOT. Scaffolds were either decellularized allo- or xenogenic biological valved conduits or bioabsorbable prosthetic materials (poly-4-hydroxybutyrate (P4HB), poly-L-lactide (PCLA), polyglycolic acid (PGA)) designed in unvalved patches,28–32 non-valved tubes,33–35 or valved tubes.36–40 Decellularized scaffolds. Dohmen et al. published an account of the first clinical implantation of a tissue-engineered heart valve in 200041: an in vitro seeded decellularized pulmonary allograft was implanted during a Ross operation in an adult patient. The 10-year clinical results of these tissue-engineered heart valves of the same group were promising despite a limited number of patients.42 Da Costa et al.43 demonstrated an excellent hemodynamic behavior and a significant decrease in human leukocyte antigen (HLA) class I and II antigens in decellularized allografts compared with standard allografts. Nevertheless pejorative clinical outcomes of this technology were also reported: Simon et al.44 showed that the Synergraft technology failed in four grafts after 2 days and 1 year post-implantation and that no recellularization of the decellularized grafts was seen at up to 1 year of follow-up. In 2010, Da Costa et al.45 investigated the outcomes of decellularized aortic homograft implants as an aortic root replacement in 41 patients. No reoperations were performed due to aortic valve dysfunction with a maximal follow-up of 53 months. Polymer scaffolds and in situ regeneration concept. The literature reports that polymer scaffolds were seeded (or not) with different types of autologous cells: endothelial cells, fibroblasts, myofibroblasts derived from peripheral vessels,28,32–35,36,37,39 smooth muscle cells derived from aorta or cardiomyocytes.29In vitro and in vivo studies (goats or adult syngenic rats) of these materials implanted in the RVOT demonstrated the biodegradation of the material,28,29 the endothelialization of the surface of the material,30,37,38 the synthesis of an extracellular matrix,28,33,35,37,38,46 the absence of thrombus or stenosis,36 and a low risk of calcification. In 2006, Hoerstrup et al. proved, in a pioneering work, the growth potential of a bioabsorbable non-valved tube seeded with endothelial cells and fibroblasts implanted on the pulmonary artery in a growing lamb model during 100 weeks.47 Concomitantly to this biological progress, other synthetic polymers (poly-L-lactic acid (PLLA),48 poly(epsilon-caprolactone) (PCL),49 poly(styrene-block-isobutylene-block-styrene) (SIBS),50 poly(glycerol-sebacate) (PGS)51), and other biological materials (fibrin,52 collagen,53 3D cardiac extracellular matrix,54 or hybrid materials55,56) were investigated to create tissue-engineered scaffolds for heart valves. Some polymeric matrices were made “bioactive” through the implantation of growth factors on their surface (transforming growth factor beta, bone morphogenetic protein, and vascular endothelial growth factor).57,58 Other research groups investigated strategies of “homing” and immobilization of circulating host-derived cells.59 Materials designed for RVOT reconstruction by tissue engineering using stem cells were first evaluated in vitro.60 They were bioabsorbable non-valved patches or valved tubes (PGA+/- P4HB or PGA+PLLA). The first stem cells used were human bone-marrow cells that displayed a myofibroblastic differentiation and synthetized an extracellular matrix.61 In 2007, autologous peripheral blood-derived endothelial progenitor cells and autologous bone-marrow-derived marrow stromal cells (MSC) were seeded on a bioabsorbable non-valved patch on the pulmonary artery of seven goats with a follow-up of 6 weeks.62–66 This study showed the development of a living and organized tissue, integrated to the native pulmonary artery. The use of bioreactors for cell culture and maturation in dynamic conditions allowed for the maturation of the tissue-engineered device, the in vitro cell differentiation, and the formation of the extracellular matrix.67–72 A non-invasive percutaneous method of implantation of tissue-engineered heart valves was described by Dr Hoerstrup’s group73 and by Emmert et al.74 From 2002, the cells used have been derived from human umbilical cord blood, Wharton’s jelly, amniotic liquid, chorial villosities, or induced pluripotent cells seeded on non-valved patches or valved tubes.75–83 Even periodontal ligament cells cultured under steady flow environments demonstrated potential for use in heart valve tissue engineering. Materials made of co-polymer of poly(lactic acid) (PLA) and polycaprolactone (PCL), seeded with human bone-marrow cells, were implanted by Shin’oka et al. in 42 patients with congenital heart diseases in Japan between 2001 and 2005.84,85 The incidence of early stenosis led this group to go back “from bed to bench” to further understand the mechanisms of this type of early failure.86 Prototypes of a bioabsorbable valve and valved tube created using PLLA reinforced with non-absorbable polyester (PET) were assessed as tissue-engineered devices to reconstruct the RVOT by the group of Menasché and Kalfa et al. (Figure 1).  | Figure 1 A Global View of a Bioabsorbable Valve Made of Poly-L-lactic acid (PLLA) and Polyester (PET). |
Table 2 summarizes the different types of synthetic polymers used in the research field of the right ventricular outflow tract.  | Table 2 Different Types of Synthetic Polymers Used in the Research Field of the Right Ventricular Outflow Tract. |
The concept of decellularization of tissue-engineered heart valves, initially made of bio-degradable synthetic materials and homologous cells, was then introduced to offer an alternative starter matrix for guided tissue regeneration.107 This decellularization phase of tissue-engineered heart valves was demonstrated not to alter the collagen structure or tissue strength; it also favored valve performance when compared to their cell-populated counterparts and could provide largely available off-the-shelf homologous scaffolds suitable for reseeding with autologous cells. Key requirements and properties of those substrates were then discussed in the light of current trends toward designing biologically inspired microenvironments for in situ tissue engineering purposes.108 The concept of in situ tissue engineering, i.e. neotissue regeneration without the use of seeded cells, could solve the disadvantages of using any cell source and achieve a versatile and easier cell-free protocol.109 The evaluation of in situ tissue engineering vasculature (iTEV) by implantation of scaffolds made of polyglycolide knitted fibers and an L-lactide and ɛ-caprolactone co-polymer sponge in the inferior vena cava of a canine model supported this concept by demonstrating a native tissue-like histological regeneration, with acceptable biomechanical characteristics.110 More recently, hundreds of polymers were comprehensively assessed for tissue engineering of cardiac valves, using polymer microarray technology.111 Biomechanical tests with real-time displacement and strain mapping were also recently reported to quantify biomechanical and biochemical properties of semilunar heart valve tissues, and potentially facilitate the development of tissue-engineered heart valves.112 The role of substrate stiffness in modulating the gene expression and phenotype of neonatal cardiomyocytes in vitro113 or seeded human bone-marrow stem cells,114 on the one hand, and in modulating the activation of valvular interstitial cells,115 on the other hand, demonstrated the importance of the mechanical properties of materials used for valve repair or for engineering valve tissue.116 Electrospinning appears in the literature as a promising technology to produce scaffolds for cardiovascular tissue engineering. Amoroso et al. evaluated the effect of processing variables and secondary fiber populations on the microstructure and the tensile and bending mechanics of electro-spun biodegradable polyurethane scaffolds for heart valve tissue engineering.117 Computational tools were developed in order to describe and predict the mechanical behavior of electrospun valve-shaped scaffolds characterized by different microstructures and showed that a pronounced degree of anisotropy was necessary to reproduce the deformation patterns observed in the native heart valve.118 In the emerging field of tissue engineering and regenerative medicine, different design strategies were evaluated to promote the development and evaluation of improved tissue engineering scaffolds. These include mimicking the extracellular matrix, predicting the structural architecture, ensuring adequate initial mechanical integrity, modifying the surface chemistry109,110,119 and topography120 to provide cell signaling, and anticipating the material selection so as to predict the required rate of bioresorption.121 The biofunctionalization of polymeric scaffolds or decellularized native homografts with motifs (such as RGD, SDF-1α, fibronectin, collagen, CD33) led to encouraging results and could be an alternative way to the complex techniques of cell culture and cell seeding.109,110,122 Prokoph et al. demonstrated that sustained delivery of SDF-1α from proangiogenic hydrogels could effectively attract early endothelial progenitor cells (ePCs), offering a powerful means to trigger endogenous mechanisms of cardiac regeneration.122 |
REFERENCES 1. Warnes CA, Williams RG, Bashore TM, et al. ACC/AHA 2008 Guidelines for the Management of Adults with Congenital Heart Disease: a report of the American College of Cardiology/American Heart Association. Task Force on Practice Guidelines (writing committee to develop guidelines on the management of adults with congenital heart disease). Circulation. 2008;118:e714–833. http://dx.doi.org/10.1161/CIRCULATIONAHA.108.190690. 2. Robotin-Johnson MC, Swanson PE, Johnson DC, Schuessler RB, Cox JL. An experimental model of small intestinal submucosa as a growing vascular graft. J Thorac Cardiovasc Surg. 1998;116:805–11. http://dx.doi.org/10.1016/S0022-5223(98)00436-X. 3. Ruiz CE, Iemura M, Medie S, et al. Transcatheter placement of a low-profile biodegradable pulmonary valve made of small intestinal submucosa: a long-term study in a swine model. J Thorac Cardiovasc Surg. 2005;130:477–84. http://dx.doi.org/10.1016/j.jtcvs.2005.04.008. 4. Wainwright JM, Hashizume R, Fujimoto KL, et al. Right ventricular outflow tract repair with a cardiac biologic scaffold. Cells Tissues Organs. 2012;195:159–70. http://dx.doi.org/10.1159/000331400. 5. Scholl FB, Boucek MM, Chan K-C, Valdes-Cruz L, Perryman R. Preliminary experience with cardiac reconstruction using decellularized porcine extracellular matrix scaffold: human applications in congenital heart disease. World J Pediatr Congenital Heart Surg. 2010;1:132–6. http://dx.doi.org/10.1177/2150135110362092. 6. Quarti A, Nardone S, Colaneri M, Santoro G, Pozzi M. Preliminary experience in the use of an extracellular matrix to repair congenital heart diseases. Interact Cardiovasc Thorac Surg. 2011;13:569–72. http://dx.doi.org/10.1510/icvts.2011.280016. 7. Witt RG, Raff G, Van Gundy J, Rodgers-Ohlau M, Si MS. Short-term experience of porcine small intestinal submucosa patches in paediatric cardiovascular surgery. Eur J Cardiothorac Surg. 2013;44:72–6. http://dx.doi.org/10.1093/ejcts/ezs638. 8. Badylak S, Kokini K, Tullius B, Simmons-Byrd A, Morff R. Morphologic study of small intestinal submucosa as a body wall repair device. J Surg Res. 2002;103:190–202. http://dx.doi.org/10.1006/jsre.2001.6349. 9. Frigiola A, Redington AN, Cullen S, Vogel M. Pulmonary regurgitation is an important determinant of right ventricular contractile dysfunction in patients with surgically repaired tetralogy of Fallot. Circulation. 2004;110(11 Suppl 1):II153–7. 10. Bonhoeffer P, Boudjemline Y, Saliba Z, et al. Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction. Lancet. 2000;356:1403–5. http://dx.doi.org/10.1016/S0140-6736(00)02844-0. 14. Nordmeyer J, Khambadkone S, Coats L, et al. Risk stratification, systematic classification, and anticipatory management strategies for stent fracture after percutaneous pulmonary valve implantation. Circulation. 2007;115:1392–7. http://dx.doi.org/10.1161/CIRCULATIONAHA.106.674259. 15. Chowdhury SM, Hijazi ZM, Rhodes J, et al. Early echocardiographic changes after percutaneous implantation of the Edwards SAPIEN transcatheter heart valve in the pulmonary position Echocardiography 2013 Feb 22; . [Epub ahead of print] http://dx.doi.org/10.1111/echo.12147. 16. Kalfa D, Loundou A, Nouaille de Gorce Y, et al. Pulmonary position cryopreserved homograft in non-Ross patients: how to improve the results? Eur J Cardiothorac Surg. 2012;42:981–7. http://dx.doi.org/10.1093/ejcts/ezs248. 17. Kalfa D, Feier H, Loundou A, et al. Cryopreserved homograft in the Ross procedure: outcomes and prognostic factors. J Heart Valve Dis. 2011;20:571–81. 19. Kwak JG, Lee JR, Kim WH, Kim YJ. Mid-term results of the Hancock II valve and Carpentier-Edward Perimount valve in the pulmonary portion in congenital heart disease. Heart Lung Circ. 2010;19:243–6. http://dx.doi.org/10.1016/j.hlc.2009.10.009. 20. Brown JW, Ruzmetov M, Rodefeld MD, Vijay P, Turrentine MW. Right ventricular outflow tract reconstruction with an allograft conduit in non-ross patients: risk factors for allograft dysfunction and failure. Ann Thorac Surg. 2005;80:655–63. http://dx.doi.org/10.1016/j.athoracsur.2005.02.053. 21. Yan SM, Mishaly D, Shinfeld A, Raanani E. Right ventricular outflow tract reconstruction: valved conduit of choice and clinical outcomes. J Cardiovasc Med (Hagerstown). 2008;9:327–37. http://dx.doi.org/10.2459/JCM.0b013e32821626ce. 24. Urso S, Rega F, Meuris B, et al. The Contegra conduit in the right ventricular outflow tract is an independent risk factor for graft replacement. Eur J Cardiothorac Surg. 2011;40:603–9. 25. Holmes AA, Co S, Human DG, Leblanc JG, Campbell AI. The Contegra conduit: late outcomes in right ventricular outflow tract reconstruction. Ann Pediatr Cardiol. 2012;5:27–33. http://dx.doi.org/10.4103/0974-2069.93706. 28. Stock UA, Sakamoto T, Hatsuoka S, et al. Patch augmentation of the pulmonary artery with bioabsorbable polymers and autologous cell seeding. J Thorac Cardiovasc Surg. 2000;120:1158–67. http://dx.doi.org/10.1067/mtc.2000.109539. 29. Sakai T, Li RK, Weisel RD, et al. The fate of a tissue-engineered cardiac graft in the right ventricular out-flow tract of the rat. J Thorac Cardiovasc Surg. 2001;121:932–42. http://dx.doi.org/10.1067/mtc.2001.113600. 30. Ozawa T, Mickle DA, Weisel RD, Koyama N, Ozawa S, Li RK. Optimal biomaterial for creation of autologous cardiac grafts. Circulation. 2002;106(12 Suppl 1):I176–82. 31. Ozawa T, Mickle DA, Weisel RD, et al. Histologic changes of nonbiodegradable and biodegradable bio-materials used to repair right ventricular heart defects in rats. J Thorac Cardiovasc Surg. 2002;124:1157–64. http://dx.doi.org/10.1067/mtc.2002.127449. 32. Ozawa T, Mickle DA, Weisel RD, et al. Tissue-engineered grafts matured in the right ventricular outflow tract. Cell Transplant. 2004;1:169–77. 35. Leyh RG, Wilhelmi M, Rebe P, Ciboutari S, Haverich A, Mertsching H. Tissue engineering of viable pulmonary arteries for surgical correction of congenital heart defects. Ann Thorac Surg. 2006;81:1466–70. http://dx.doi.org/10.1016/j.athoracsur.2005.08.083. 37. Steinhoff G, Stock U, Karim N, et al. Tissue engineering of pulmonary heart valves on allogenic acellular matrix conduits: in vivo restoration of valve tissue. Circulation. 102(19 Suppl 3):III50–5. 200. 39. Leyh R, Wilhelmi M, Haverich A, Mertsching H. [A xenogeneic acellularized matrix for heart valve tissue engineering: in vivo study in a sheep model]. Z Kardiol. 2003;92:938–46. [Article in German] http://dx.doi.org/10.1007/s00392-003-0978-4. 43. da Costa FD, Dohmen PM, Duarte D, et al. Immunological and echocardiographic evaluation of decellularized versus cryopreserved allografts during the Ross operation. Eur J Cardiothorac Surg. 2005;27:572–8. http://dx.doi.org/10.1016/j.ejcts.2004.12.057. 46. Sales VL, Engelmayr GC Jr, Johnson JA Jr, et al. Protein precoating of elastomeric tissue-engineering scaffolds increased cellularity, enhanced extracellular matrix protein production, and differentially regulated the phenotypes of circulating endothelial progenitor cells. Circulation. 2007;116(11 Suppl):I55–6. 47. Hoerstrup SP, Cummings Mrcs I, Lachat M, et al. Functional growth in tissue-engineered living, vascular grafts: follow-up at 100 weeks in a large animal model. Circulation. 2006;114(1 Suppl):I159–66. 48. Eckert CE, Mikulis BT, Gottlieb D, et al. Three-dimensional quantitative micromorphology of pre and post-implanted engineered heart valve tissues. Ann Biomed Eng. 2011;39:205–22. http://dx.doi.org/10.1007/s10439-010-0162-8. 49. Del Gaudio C, Grigioni M, Bianco A, De Angelis G. Electrospun bioresorbable heart valve scaffold for tissue engineering. Int J Artif Organs. 2008;31:68–75. 50. Wang Q, McGoron AJ, Bianco R, Kato Y, Pinchuk L, Schoephoerster RT. In-vivo assessment of a novel polymer (SIBS) trileaflet heart valve. J Heart Valve Dis. 2010;19:499–505. 51. Masoumi N, Jean A, Zugates JT, Johnson KL, Engelmayr GC Jr. Laser microfabricated poly(glycerol sebacate) scaffolds for heart valve tissue engineering. J Biomed Mater Res A. 2013;101:104–14. http://dx.doi.org/10.1002/jbm.a.34305. 52. Flanagan TC, Sachweh JS, Frese J, et al. In vivo remodeling and structural characterization of fibrin-based tissue-engineered heart valves in the adult sheep model. Tissue Eng Part A. 2009;15:2965–76. http://dx.doi.org/10.1089/ten.tea.2009.0018. 53. Tedder ME, Simionescu A, Chen J, Liao J, Simionescu DT. Assembly and testing of stem cell-seeded layered collagen constructs for heart valve tissue engineering. Tissue Eng Part A. 2011;17:25–36. http://dx.doi.org/10.1089/ten.tea.2010.0138. 54. Mainwright JM, Hashizume R, Fujimoto KL, et al. Right ventricular outflow tract repair with a cardiac biologic scaffold. Cells Tissues Organs. 2012;195:159–70. http://dx.doi.org/10.1159/000331400. 56. Duan B, Wu L, Yuan X, et al. Hybrid nanofibrous membranes of PLGA/chitosan fabricated via an electrospinning array. J Biomed Mater Res A. 2007;83:868–78. http://dx.doi.org/10.1002/jbm.a.31408. 57. Chiu YN, Norris RA, Mahler G, Recknagel A, Butcher JT. Transforming growth factor β, bone morpho-genetic protein, and vascular endothelial growth factor mediate phenotype maturation and tissue remodeling by embryonic valve progenitor cells: relevance for heart valve tissue engineering. Tissue Eng Part A. 2010;16:3375–83. http://dx.doi.org/10.1089/ten.tea.2010.0027. 59. Schleicher M, Wendel HP, Fritze O, Stock UA. In vivo tissue engineering of heart valves: evolution of a novel concept. Regen Med. 2009;4:613–19. http://dx.doi.org/10.2217/rme.09.22. 63. Sales VL, Mettler BA, Lopez-Ilasaca M, Johnson JA Jr, Mayer JE Jr. Endothelial progenitor and mesenchymal stem cell-derived cells persist in tissue-engineered patch in vivo: application of green and red fluorescent protein-expressing retroviral vector. Tissue Eng. 2007;13:525–35. http://dx.doi.org/10.1089/ten.2006.0128. 65. Sales VL, Mettler BA, Engelmayr GC Jr, et al. Endothelial progenitor cells as a sole source for ex vivo seeding of tissue-engineered heart valves. Tissue Eng Part A. 2010;16:257–67. http://dx.doi.org/10.1089/ten.tea.2009.0424. 67. Van Vlimmeren MA, Driessen-Mol A, Oomens CW, Baaijens FP. The potential of prolonged tissue culture to reduce stress generation and retraction in engineered heart valve tissues. Tissue Eng Part C Methods. 2013;19:205–15. http://dx.doi.org/10.1089/ten.tec.2012.0100. 68. Rubbens MP, Driessen-Mol A, Boerboom RA, et al. Quantification of the temporal evolution of collagen orientation in mechanically conditioned engineered cardiovascular tissues. Ann Biomed Eng. 2009;37:1263–72. http://dx.doi.org/10.1007/s10439-009-9698-x. 69. Ramaswamy S, Gottlieb D, Engelmayr GC Jr, et al. The role of organ level conditioning on the promotion of engineered heart valve tissue development in-vitro using mesenchymal stem cells. Biomaterials. 2010;31:1114–25. http://dx.doi.org/10.1016/j.biomaterials.2009.10.019. 70. Martinez C, Rath S, Van Gulden S, et al. Periodontal ligament cells cultured under steady-flow environments demonstrate potential for use in heart valve tissue engineering. Tissue Eng Part A. 2013;19:458–66. http://dx.doi.org/10.1089/ten.tea.2012.0149. 73. Schmidt D, Dijkman PE, Driessen-Mol A, et al. Minimally-invasive implantation of living tissue engineered heart valves: a comprehensive approach from autologous vascular cells to stem cells. J Am Coll Cardiol. 2010;56:510–20. http://dx.doi.org/10.1016/j.jacc.2010.04.024. 74. Emmert MY, Weber B, Behr L, et al. Transapical aortic implantation of autologous marrow stromal cell-based tissue-engineered heart valves: first experiences in the systemic circulation. JACC Cardiovasc Interv. 2011;4:822–3. http://dx.doi.org/10.1016/j.jcin.2011.02.020. 77. Schmidt D, Achermann J, Odermatt B, et al. Prenatally fabricated autologous human living heart valves based on amniotic fluid derived progenitor cells as single cell source. Circulation. 2007;116(11 Suppl):I64–70. 78. Schmidt D, Mol A, Breymann C, et al. Living autologous heart valves engineered from human prenatally harvested progenitors. Circulation. 2006;114(1 Suppl):I125–31. 80. Sha JM, Yan ZY, Cheng GC, et al. In-vitro seeding of human umbilical cord vein endothelial cells on hydroxyapatite for mechanical heart valve applications. J Heart Valve Dis. 2010;19:506–12. 83. Schmidt D, Mol A, Neuenschwander S, et al. Living patches engineered from human umbilical cord derived fibroblasts and endothelial progenitor cells. Eur J Cardiothorac Surg. 2005;27:795–800. http://dx.doi.org/10.1016/j.ejcts.2005.01.064. 84. Shin’oka T, Matsumura G, Hibino N, et al. Midterm clinical result of tissue-engineered vascular autografts seeded with autologous bone marrow cells. J Thorac Cardiovasc Surg. 2005;129:1330–8. http://dx.doi.org/10.1016/j.jtcvs.2004.12.047. 88. Benton JA, Kern HB, Anseth KS. Substrate properties influence calcification in valvular interstitial cell culture. J Heart Valve Dis. 2008;17:689–99. 95. Sodian R, Sperling JS, Martin DP, et al. Fabrication of a trileaflet heart valve scaffold from a polyhydroxy-alkanoate biopolyester for use in tissue engineering. Tissue Eng. 2000;6:183–8. http://dx.doi.org/10.1089/107632700320793. 96. Schmidt D, Mol A, Odermatt B, et al. Engineering of biologically active living heart valve leaflets using human umbilical cord-derived progenitor cells. Tissue Eng. 2006;12:3223–32. http://dx.doi.org/10.1089/ten.2006.12.3223. 97. Dvorin EL, Wylie-Sears J, Kaushal S, Martin DP, Bischoff J. Quantitative evaluation of endothelial progenitors and cardiac valve endothelial cells: proliferation and differentiation on poly-glycolic acid/poly-4-hydroxybutyrate scaffold in response to vascular endothelial growth factor and transforming growth factor beta1. Tissue Eng. 2003;9:487–93. http://dx.doi.org/10.1089/107632703322066660. 98. Hoerstrup SP, Kadner A, Melnitchouk S, et al. Tissue engineering of functional trileaflet heart valves from human marrow stromal cells. Circulation. 2002;106:I143–50. 99. van Lieshout MI, Vaz CM, Rutten MC, Peters GW, Baaijens FP. Electrospinning versus knitting: two scaffolds for tissue engineering of the aortic valve. J Biomater Sci Polym Ed. 2006;17:77–89. http://dx.doi.org/10.1163/156856206774879153. 100. Sant S, Hwang CM, Lee SH, Khademhosseini A. Hybrid PGS-PCL microfibrous scaffolds with improved mechanical and biological properties. J Tissue Eng Regen Med. 2011;5:283–91. http://dx.doi.org/10.1002/term.313. 101. Courtney T, Sacks MS, Stankus J, Guan J, Wagner WR. Design and analysis of tissue engineering scaffolds that mimic soft tissue mechanical anisotropy. Biomaterials. 2006;27:3631–8. 106. Kidane AG, Burriesci G, Edirisinghe M, Ghanbari H, Bonhoeffer P, Seifalian AM. A novel nanocomposite polymer for development of synthetic heart valve leaflets. Acta Biomater. 2009;5:2409–17. http://dx.doi.org/10.1016/j.actbio.2009.02.025. 107. Dijkman PE, Driessen-Mol A, Frese L, Hoerstrup SP, Baaijens FP. Decellularized homologous tissue-engineered heart valves as off-the-shelf alternatives to xeno- and homografts. Biomaterials. 2012;33:4545–54. http://dx.doi.org/10.1016/j.biomaterials.2012.03.015. 109. Patterson JT, Gilliland T, Maxfield MW, et al. Tissue-engineered vascular grafts for use in the treatment of congenital heart disease: from the bench to the clinic and back again. Regen Med. 2012;7:409–19. http://dx.doi.org/10.2217/rme.12.12. 110. Matsumura G, Nitta N, Matsuda S, et al. Long-term results of cell-free biodegradable scaffolds for in situ tissue-engineering vasculature: in a canine inferior vena cava model. PLoS One. 2012;7:e35760. http://dx.doi.org/10.1371/journal.pone.0035760. 112. Huang HY, Balhouse BN, Huang S. Application of simple biomechanical and biochemical tests to heart valve leaflets: implications for heart valve characterization and tissue engineering. Proc Inst Mech Eng H. 2012;226:868–76. http://dx.doi.org/10.1177/0954411912455004. 114. Emani S, Mayer JE Jr, Emani SM. Gene regulation of extracellular matrix remodeling in human bone marrow stem cell-seeded tissue-engineered grafts. Tissue Eng Part A. 2011;17:2379–88. http://dx.doi.org/10.1089/ten.tea.2010.0628. 115. Quinlan AM, Billiar KL. Investigating the role of substrate stiffness in the persistence of valvular inter-stitial cell activation. J Biomed Mater Res A. 2012;100:2474–82. 116. van Vlimmeren MA, Driessen-Mol A, Oomens CW, Baaijens FP. Passive and active contributions to generated force and retraction in heart valve tissue engineering. Biomech Model Mechanobiol. 2012;11:1015–27. http://dx.doi.org/10.1007/s10237-011-0370-7. 117. Amoroso NJ, D’Amore A, Hong Y, Rivera CP, Sacks MS, Wagner WR. Microstructural manipulation of electrospun scaffolds for specific bending stiffness for heart valve tissue engineering. Acta Biomater. 2012;8:4268–77. http://dx.doi.org/10.1016/j.actbio.2012.08.002. 119. Baumann L, Prokoph S, Gabriel C, Freudenberg U, Werner C, Beck-Sickinger AG. A novel, biased-like SDF-1 derivative acts synergistically with starPEG-based heparin hydrogels and improves eEPC migration in vitro. J Control Release. 2012;162:68–75. http://dx.doi.org/10.1016/j.jconrel.2012.04.049. 124. Kohl T, Sharland G, Allan LD, et al. World experience of percutaneous ultrasound-guided balloon valvuloplasty in human fetuses with severe aortic valve obstruction. Am J Cardiol. 2000;85:1230–3. http://dx.doi.org/10.1016/S0002-9149(00)00733-5. 125. Marshall AC, van der Velde ME, Tworetzky W, et al. Creation of an atrial septal defect in utero for fetuses with hypoplastic left heart syndrome and intact or highly restrictive atrial septum. Circulation. 2004;110:253–8. http://dx.doi.org/10.1161/01.CIR.0000135471.17922.17. 127. Le Bret E, Papadatos S, Folliguet T, et al. Interruption of patent ductus arteriosus in children: robotically assisted versus videothoracoscopic surgery. J Thorac Cardiovasc Surg. 2002;123:973–6. http://dx.doi.org/10.1067/mtc.2002.121049. 132. Butler E, Folk C, Cohen A, et al. Metal MEMS tools for beating-heart tissue approximation. IEEE Int Conf Robot Autom. 2011;2011:411–16. 133. Vasilyev NV, Melnychenko I, Kitahori K, et al. Beating-heart patch closure of muscular ventricular septal defects under real-time three-dimensional echocardiographic guidance: a preclinical study. J Thorac Cardiovasc Surg. 2008;135:603–9. http://dx.doi.org/10.1016/j.jtcvs.2007.09.045. 136. Cannon JW, Howe RD, Dupont PE, Triedman JK, Marx GR, del Nido PJ. Application of robotics in congenital cardiac surgery. Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2003;6:72–83. http://dx.doi.org/10.1053/pcsu.2003.50000. |